Chapter 5. TESTING AND FAULT FINDING
60 | ADTS 405 Mk2 Instruction Manual-English
© 2015 General Electric Company – All rights reserved.
If the display shows ERROR 703 START UP Ps SELF TEST TIME ERROR. Check that the expansion port
plug (item 13, Table 2-1) is correctly connected in the expansion port on the rear panel (Fig. 2-4).
Note:
Some errors display extra information using a four digit hexadecimal
code. This should always be recorded with the error number to provide
information for the repair depot.
5.3 Standard Serviceability Test
The following procedure can be used to check the functions and facilities of the ADTS 405. For
further information refer to/see Section 6, “REFERENCE AND SPECIFICATION”.
In the following procedures:
• All key presses are highlighted in
bold
and shown as identified on the key-pad.
• Key presses inside brackets e.g, [MORE], are soft key presses (i.e, function key selections
{F1 to F4} indicated on the display screen).
Note:
When the display shows a message indicating entry of incorrect data or
values, pressing
CLEAR/QUIT
clears the message and allows the correct entry to
be made.
5.3.1 Procedure
5.3.1.1 Connection and power-on checks
a.
Connect power to the unit and, if necessary, connect the hand terminal.
b.
Make sure that the blanking caps are fitted to the Ps and Pt front panel outputs.
c.
Check power indicator is on.
d.
Set the front panel switch to OPERATE.
e.
Set the front panel switch for the pump unit to PUMP ON and check that the pump power
indicator lights.
f.
Check the display shows the first stage power-up message as below (where: xx is the
version number of the installed display driver software):
g.
Check the second stage power-up message displays as below (where: xx is the version
number of the installed main operating software):
Note:
“ADTS 405" may be replaced by an alternative identity string. The
software version may also be supplemented by extra alpha characters.
DK 425 V.xx
Display Power Up
Please Wait
DRUCK
ADTS 405
DK 415
VER.xx
Summary of Contents for Druck ADTS 405 R/F Mk2
Page 2: ......
Page 14: ...xii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 16: ...xiv ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 18: ...xvi ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 121: ......