Chapter 3. OPERATION
34 | ADTS 405 Mk2 Instruction Manual-English
© 2015 General Electric Company – All rights reserved.
3.5.6.3 Airspeed Switch test
The following example shows how an airspeed switch can be functionally checked.
Note:
For low airspeed switches (i.e. 130 knots) the Pt Only facility could be used.
Example
a.
Press
LEAK MEASURE/CONTROL
to enter control mode. Press
SPEED/Qc
.
b.
If necessary, press
RATE
and enter a change of airspeed low enough to observe switch
operation.
c.
Enter an Aim value below the operating limits of the switch (i.e., airspeed switch operating
at 130 knots ±2 knots set 127 knots).
d.
Press
ENTER
and observe the airspeed changing.
e.
Using the nudge facility, set to an appropriate step value, increase the Aim value.
f.
Increase the Aim until the switch operates (contacts close) and record the value.
g.
Change the Aim to above the airspeed operating range (i.e. 133 knots). Repeat the steps 3
to 5, decreasing the Aim value until the airspeed switch operates (contacts open) and
record value.
This procedure can be repeated. If many operations of a switch are required use the oscillating
facility detailed in Full Set-up, see 6.2.30. The above method (steps a. to g.) can be applied for the
other parameters:
Examples:
Altitude switch
• Low altitude warning
• Landing gear configuration warning
• Height lock
Mach switch
• Speed brake scheduling
• Mach cruise lock
3.5.7 Engine Pressure Ratio (EPR)
The ADTS 405 may be used to check EPR sensors and indicators. Use Ps for INLET pressure and Pt
for OUTLET pressure. To carry out an EPR check, the display must be showing pressure units e.g,
mbar or in Hg.
3.5.7.1 Example EPR
To enter an EPR of 1.8 with inlet pressure of 500 mbar (15 inHg), proceed as follows:
a.
If the display shows “Leak Measure”, press
LEAK MEASURE/CONTROL
to regain control.
Note:
The pressure/vacuum pumps must be switched on.
b.
Press
ROC/Ps RATE
to select rate of change of static.
c.
Enter required rate of change e.g, 1000 mbar/min, (30 inHg/min).
Summary of Contents for Druck ADTS 405 R/F Mk2
Page 2: ......
Page 14: ...xii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 16: ...xiv ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 18: ...xvi ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 121: ......