Operating Procedures
© 2015 General Electric Company – All rights reserved.
ADTS 405 Mk2 Instruction Manual-English | 27
Figure 3-6: Rate Timer Display Pressure Units
Pt Only Display
(Fig 3.7)
The Pt only display overrides the other display modes (single, dual and triple). The measured
altitude and the CAS are produced from Ps in measure mode.
Figure 3-7: Pt Only Display
3.5.1.1 Change display units:
a.
Press
SETUP
.
b.
Select [MORE].
c.
Select [DISPLAYS/OPTIONS].
d.
Select [DISPLAY TYPE].
e.
Using [NEXT] or [PREV].
f.
Select either SINGLE, DUAL or TRIPLE, ENCODER (
more display types with options fitted
).
g.
Press [SAVE] to accept.
h.
Press
CLEAR/QUIT
repeatedly to return to user display.
Parameters
Measured
Timed
values
T for timed rate
of change
Special
message
Units of
measurement
Timed Rates
Rt Ps
Rt Pt
Rt Qc
0.637 T
0.341 T
-0.224 T
mbar/m
mbar/m
mbar/m
Alt always
shown
Measured
altitude
T for timed rate
of change
ft
kts
Selected
parameter
Aim (set-point or target)
or “leak-measure” if
Pitot controller selected off
Pointer for
parameter
Special
messages
Alt
Pt Only
CAS
Aim
407
0.0
0.0
Summary of Contents for Druck ADTS 405 R/F Mk2
Page 2: ......
Page 14: ...xii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 16: ...xiv ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 18: ...xvi ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 121: ......