6.2.49 SETUP, EXECUTE TEST PROGRAM
This function sets up the results file options.
Result data generated by the test program can either be sent out to the printer port or saved to a
result file and stored in the ADTS 405. Pressing F1, the [PRINT/SAVE] key, sets the option. Any of the
results files, stored in the ADTS 405, can be sent to the printer port. Press F2, the [PRINT FILE] key,
to select the required file.
6.2.50 SETUP, HELP
This function provides help on the set-up key.
6.2.51 SETUP,
or
(Nudge keys)
This function allows the increment value of the nudge key to be adjusted for every parameter.
1.
When
SETUP
,
or
are pressed, the display shows the value of nudge for the parameter
last displayed.
2.
Enter a new value to replace the present value.
3.
Use [NEXT] or [PREV] to display other parameters.
4.
Press
CLEAR/QUIT
when the nudge values of all the required parameters have been
changed.
In aeronautical units, the parameters available are:
In pressure units, the parameters available are:
5.
The nudge value can be used for large increments e.g., 100 kts or 5000 ft to “step through”
the cardinal points of a test.
CAUTION
Make sure the nudge increment is set to a suitable value before using the
nudge facility to change the aim value.
6.2.52 MINIMUM SETUP
Minimum set-up allows access to a limited number of set-up functions. Any parameter changed
under set-up returns to the default settings at power-down. The CONFIGuration mode may be used
to change the default settings so that the ADTS 405 powers-up in the required state.
6.2.53 SETUP, [UNITS], [AERO]
The current operating limits are changed to the default aeronautical units. These units can be
programmed using CONFIG, [UNITS], [AERO].
6.2.54 SETUP, [UNITS], [PRESS]
The current operating limits are changed to the default pressure units. These units can be
programmed using CONFIG, [UNITS], [PRESS].
ALT
CAS/TAS
MACH
ROC
RATE CAS
Ps
Pt
Qc
EPR
RATE Ps
-
RATE Qc
RATE EPR
Summary of Contents for Druck ADTS 405 R/F Mk2
Page 2: ......
Page 14: ...xii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 16: ...xiv ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 18: ...xvi ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Page 121: ......