Revision A
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
25
2018554-001
1.10 TROUBLESHOOTING
Trouble
Probable Cause
Fault Isolation
Monitor will not switch on
when powered by battery
- Battery may be discharged
- PSU board fault
- Replace with known, charged battery
- Replace PSU board
Unit will not switch on from
either battery or external AC
source. No AC mains light.
- Locked processor
- PSU Module
- Press and hold ON/OFF button for 15
seconds.
- check PSU module for 24 Vdc
Pressing rotor fails to power
on the unit, menu selection is
possible
Rotor ON/OFF switch
Replace mains PWA
No rotor selection or menu
skips
Main board failure
Replace main board
Illegible or no LCD display
- LCD failure
- Main board fault (usually
accompanied by an audible
alarm if main board processor
has ceased operation)
- Negative supply to LCD (-9.6
Vdc) is out of tolerance
- Replace LCD
- Replace main board
- Check negative supply reaching the
LCD module
No LCD backlight
- LCD faulty
- Main board fault
- Replace the LCD
- Replace the main board
Incorrect, missing segments,
or no LED display output
- Main board fault (usually
accompanied by an audible
alarm if main board processor
has ceased operation)
- No or low +4 Vdc LED
supply
- Replace LCD
- Replace PSU board
- Check supply continuity from PSU board
to Main board
Battery not charging or not
holding charge
- Check battery charging
function
- PSU PWA
- Input Service code: 8378 after selecting
the Service button. Then select
PIC
. Then
select
C bat
. The Monitor displays the
Battery C value
. Multiply Battery C value
(ADU) x .101 = charge volts. If charge
volts is less than 13.0 volts, replace PSU
PWA
“SpO2 probe disconnected”
error displayed on LCD
- Faulty probe
- SpO2 PWA failed
- Test parameter with known good
accessories
- Replace PWA and test unit with Spo2
simulator
Summary of Contents for Dinamap PRO Series
Page 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410...
Page 2: ......
Page 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Page 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Page 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix...
Page 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001...
Page 41: ......