4
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
Revision A
2018554-001
Avoid storing batteries at temperatures above 77° F (25° C). High
storage temperatures can dramatically increase the self-
discharge rate of battery.
1.3.5 Checking Calibration
Perform the test procedures described in 4.6 whenever the accuracy
of any of the parameters is in doubt.
1.4 Clinician Mode
The PRO Monitor provides a special mode named Clinician Mode.
This mode is accessed from the
More…
option on the Main menu,
and requires a code before a user can access it.
The
Clinician Mode
enables the user to:
Verify NIBP calibration and PM dates
Change the alarms to “Permanent Silence” mode
Set target inflation pressure
1.4.1 Accessing the Clinician Mode
This section provides an overview of the Clinician menus. For more
information and details of operation, see the
PRO Series 100-400
Monitor Operation Manual.
To enter Clinician Mode from the Main menu, use the rotor control
and LCD.
1. Select the
More…
button.
2. Select the
Service
button.
Summary of Contents for Dinamap PRO Series
Page 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410...
Page 2: ......
Page 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Page 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Page 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix...
Page 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001...
Page 41: ......