14
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
Revision A
2018554-001
off. The
Save
button must then be selected for this to be a permanent
change.
Cancel
Selection of this button takes the user back to the
Service
Mode
System Menu. Changes to the Asset ID tag will
not
be retained if the
Save
button has not been selected before leaving the menu.
Save
Selection of this button enters the Asset ID Tag number being created
in eeprom. None of the operations of the
Digit, Clear,
or
Back
buttons are permanent until the
Save
button is selected. On power-
up, this saved number will be displayed on start-up screen.
1.6 PRE-SERVICE TEST PROCEDURES
The following procedures can be used to check the primary functions
of the PRO Monitor before releasing the unit for clinical service.
The Monitor is tested by applying various stimuli to the sensor
interfaces or by measuring of specific parameters. The test
procedures employ features of the operational software and test
modes of the service menus.
A guide to the Monitor controls, indicators, and connectors is in
Section 2 of the PRO Monitor service manual.
Test Equipment
1. Digital Pressure Gauge 0-375mmHg, 0.2% accuracy, Digitron
P200L or similar.
2. NELLCOR
®
SpO
2
Simulator Model SRC-MAX or similar for Nellcor
SpO
2
or equivalent.
3. Biotek SpO
2
Simulator for Masimo
®
SpO
2
or equivalent.
4. NIBP Calibration Kit P/N 320246, or similar.
5. IVAC
®
Probe Simulator TE 1811.
Test Conditions
Testing shall be conducted with an ambient temperature of 25
°
C
±
5
°
C (77 °F
±
9 °F).
1.6.1 SpO
2
Checks (PRO 300 & 400)
For Monitors equipped with Nellcor SpO
2
On occasion when testing the integrity of the Nellcor oximetry system,
abnormal results may occur when introducing large changes in the
pulse rate and/or pulse amplitude. Extreme changes in rate sent to
the Nellcor sensor by the SpO
2
simulator may cause the SpO
2
algorithm to completely miss finding the pulse rate.
Summary of Contents for Dinamap PRO Series
Page 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410...
Page 2: ......
Page 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Page 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Page 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix...
Page 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001...
Page 41: ......