Revision A
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
9
2018554-001
1.5.2 Service Mode Main Menu
The options accessed through the Main Service menu, pictured
above, allow the technician to check basic functions of the PRO
Monitor, calibrate the NIBP system, and set the language of
operation.
Note:
These menus and adjustments should be used only by
qualified service technicians.
NIBP Button
The NIBP button accesses the NIBP Service menu. Use this menu to
calibrate and check the NIBP transducers and the overpressure
circuitry (Section 4.5.3).
Info Button
This button displays the date of the most recent NIBP calibration and
PM check. Select
OK
to return the display to the Service menu.
System Button
This button accesses the System menu. The options on this menu set
the display language, communications protocols, test the EEPROM,
and print the error log (Section 4.5.4).
Silence Button
Selecting the
Silence
button displays a request to confirm the choice.
This is a working option available only through the Clinician Mode
(1234). Alarms are disabled when in Service Mode.
Yes
mutes
all
patient
alarms until the PRO Monitor is powered off and
on again or the alarm Silence button is selected again to enable the
alarms. A confirmation menu appears (pictured above) on the display.
Selecting either
Yes
or
No
exits the menu. If silence is confirmed, the
alarm Silence button on the front panel illuminates momentarily.
Summary of Contents for Dinamap PRO Series
Page 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410...
Page 2: ......
Page 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Page 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Page 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix...
Page 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001...
Page 41: ......