Revision A
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
13
2018554-001
The
Address
box sets the Monitor address within a range of 32 to
126. Address 32 is not unique. All monitors, regardless of unit
address, respond to commands to unit address 32.
Selecting the
1846
button sets the comms protocol to 1846 format,
and the baud rate to 600 bps.
The baud
Rate
can be set to
300, 600, 1200, 2400, 4800, 9600,
or
19200
.
Select
OK
to return to the Main Service menu.
eprom Button (EEPROM Testing)
This button initiates a test of the read and write functions on the
EEPROM. The test is repeated 20 times. The results are displayed on
the LCD and recorded in the error log.
Err Log Button
This button generates a printout of the last 20 system errors that have
occurred on the unit. The system and secondary processor software
versions are also listed on the error log.
ID Tag Button
Selection of the
ID Tag
button opens a menu that allows the user to
enter a 12-digit Monitor Asset Tag for their own use. If the Asset ID
Tag is not equal to zero, the ID tag number will be displayed at the top
of the screen upon power-up, before the Monitor displays the Main
Menu. If the value is equal to zero, the ID Tag option is disabled and
not shown on the start-up screen upon power-up.
Input Asset Tag Digits
Selection of the individual digit buttons (
0
to
9
) in this menu allows the
user to input an Asset ID Tag. Use the
Select
Knob to choose the
digits of the tag for the Monitor. You must select
Save
to permanently
change the Asset ID Tag. The choice of digits range from
0
to
9
.
Back
Selection of this button erases the last digit chosen for the Asset ID
tag. The
Save
button must then be selected for this to be a
permanent change.
Clear
Selection of this button resets the Asset ID tag number being created
back to zero, allowing a new number to be input or to turn the feature
Summary of Contents for Dinamap PRO Series
Page 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410...
Page 2: ......
Page 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Page 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Page 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix...
Page 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001...
Page 41: ......