– 60 –
VOLTAGE
RESISTANCE
TEMPERATURE
.95
20066
50°F
1.56
10450
75°F
2.25
5824
100°F
2.91
3411
125°F
3.48
2081
150°F
3.91
1330
175°F
Thermistor
The Thermistor is located inside the turbidity sensor.
The thermistor monitors water temperature in the
tub. If the thermistor opens or shorts it will cause an
error (see Consumer Error Mode). If it fails in open or
shorted, the control will default to the longest time
algorithm.
7KHUPLVWRU6SHFL¿FDWLRQV
Turbidity Sensor Calibration
&DOLEUDWLRQRFFXUVHYHU\F\FOHV$IWHUWKH¿QDO
rinse but before dry, the control will add three extra
ULQVHF\FOHV7KH¿UVWF\FOHZLOO¿OOFLUFXODWHWZR
PLQXWHVDQGGUDLQ7KHVHFRQGF\FOHZLOOEHD¿OODQG
GUDLQ7KHWKLUGF\FOHZLOOEHD¿OOFLUFXODWHFDOLEUDWH
and drain. The cycle will now advance to dry and
complete.
Replacement controls will enter a calibration at the
HQGRIWKH¿UVWF\FOHWKDWWKHFRQWUROFRPSOHWHVDV
PHQWLRQHGDERYH,IWKH¿UVWF\FOHLVLQWHUUXSWHGDQG
calibration does not occur, calibration will retry on
the next cycle, until it is completed. Calibration will
not occur on demand.
Turbidity Sensor Diagnostics
If the turbidity sensor fails open or shorted, the
FRQWUROZLOOGHIDXOWWRWKHPD[LPXP¿OOVDQG
circulation time that the control is programmed for.
Use Diagnostic Fault Code Display Mode to look for
Turbidity Sensor Faults.
To check the thermistor, access the Main Control
(see To Access Main Control Board section), check
resistance on connector J711, pin 7, YR to pin 10,
WG. The sensor may also be removed and the two
outside terminals may be checked. Use the table to
the left to calculate the correct resistance reading.
Turbidity Sensor Removal
To remove the sensor, empty the sump of water.
Disconnect the wire harness, turn the sensor
counter-clockwise and pull from the sump. To
reinstall, push in and turn clockwise.