5-8
170 Series Monitor
Revision D
2003023-001
Fetal Heart Rate Monitoring:
Heartbeat Coincidence
Strip Chart Annotation
When heartbeat coincidence detection is enabled, and both heart rate channels are
active, the annotation
HBC
prints in the center margin
of the strip chart paper following the active FHR modes. (Refer to
.)
As soon as heartbeat coincidence is detected, two overlaid hearts
print in the
upper portion of the top grid of the strip chart paper; the hearts print twice per page
for as long as coincidence is detected. Once coincidence is resolved, two side-by-
side hearts
print once. (Refer to
.)
If you disconnect a transducer while coincidence is detected, the overlaid hearts
stop printing and the side-by-side hearts
print once. In addition, the mode status
line prints on the strip chart paper—without the
HBC
annotation—indicating the
deactivated mode.
Figure 5-3. Simulated Heartbeat Coincidence Detection Trace
Coincidence detected.
Coincidence resolved.
Coincidence enabled.
This annotation prints
every ten minutes along
with the modes.
Simulated
trends.
Summary of Contents for Corometrics 170 Series
Page 1: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Page 2: ......
Page 3: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Page 12: ......
Page 21: ...Revision D 170 Series Monitor 1 9 2003023 001 Safety Definitions of Symbols For your notes ...
Page 22: ...1 10 170 Series Monitor Revision D 2003023 001 Safety Definitions of Symbols ...
Page 42: ...3 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 58: ...4 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 78: ...6 6 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 106: ...10 4 170 Series Monitor Revision D 2003023 001 Supplies Accessories ...
Page 111: ......