4-10
170 Series Monitor
Revision D
2003023-001
Setup Procedures:
Customizing the Monitor
Customizing the Monitor
The monitor includes a user setup mode where you can:
enable/disable alarm functionality
set the high alarm limit for the fetal heart rate
set the low alarm limit for the fetal heart rate
set the alarm volume
set the time and date
(A 170 Series Monitor is Year 2000 compliant.)
NOTE:
For Models 172, 173, and 174, use the leftmost set of volume controls.
Figure 4-12. Setup Mode Summary (Model 172 shown)
2
1
2
1
2
1
Press
to enter the
user setup mode.
Use ( ) to select
feature code 2FHR
High Alarm Limit.
See Table 4-2.
Press to switch
displays. The
heartbeat indicator
lights.
2
1
Use ( ) to change
the value to 185
BPM. See Table
4-2.
Press to switch
displays again. The
± sign lights. (Repeat
steps 2 to 5 .)
1
3
4
5
2
Monitor Feature Code
(e.g. FHR Alarms or
High Alarm Limit).
Press to switch
between displays.
Use ( or ) to
change the number.
Setting or value for
selected feature (e.g.
FHR Alarms On or 180
BPM High Alarm Limit).
Use ( or ) to
change the number.
2
1
Press to switch between displays.
Repeat steps to change other settings.
EXAMPLE
Summary of Contents for Corometrics 170 Series
Page 1: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Page 2: ......
Page 3: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Page 12: ......
Page 21: ...Revision D 170 Series Monitor 1 9 2003023 001 Safety Definitions of Symbols For your notes ...
Page 22: ...1 10 170 Series Monitor Revision D 2003023 001 Safety Definitions of Symbols ...
Page 42: ...3 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 58: ...4 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 78: ...6 6 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 106: ...10 4 170 Series Monitor Revision D 2003023 001 Supplies Accessories ...
Page 111: ......