1-6
170 Series Monitor
Revision D
2003023-001
Safety:
Monitor Safety Information
Cautions
CAUTIONS
STATIC SENSITIVITY—This monitor is extremely static
sensitive and should be handled using electrostatic discharge
precautions.
ANNUAL SERVICING—For continued safety and performance
of the monitor, it is recommended that the calibration, accuracy,
and electrical safety of the monitor be verified on an annual basis
by a GE Medical Systems
Information Technologies
Service
Representative.
DAILY TESTING—It is essential that the monitor and
accessories be inspected every day. It is recommended practice to
ensure the monitor passes its self-test routine—initiated each time
the monitor is turned on. Refer to
ENVIRONMENT—The performance of the monitor has not been
tested in certain areas, such as x-ray and imaging suites. The
monitor is not recommended for use in these environments.
PERFORMANCE—Report all problems experienced with the
monitor. If the monitor is not working properly, contact your
Service Representative. The monitor should not be used if it is not
working properly.
Summary of Contents for Corometrics 170 Series
Page 1: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Page 2: ......
Page 3: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Page 12: ......
Page 21: ...Revision D 170 Series Monitor 1 9 2003023 001 Safety Definitions of Symbols For your notes ...
Page 22: ...1 10 170 Series Monitor Revision D 2003023 001 Safety Definitions of Symbols ...
Page 42: ...3 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 58: ...4 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 78: ...6 6 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 106: ...10 4 170 Series Monitor Revision D 2003023 001 Supplies Accessories ...
Page 111: ......