Revision D
170 Series Monitor
4-13
2003023-001
Setup Procedures:
Customizing the Monitor
* = service setup mode
Table 4-3. Summary of Factory Defaults
Setup Option
Factory Default
Hospital/Clinic Setting
FHR Alarms
on
FHR High Alarm Limit
160 BPM
FHR Low Alarm Limit
120 BPM
FHR Alarm Volume
5
Time/Date
Eastern Standard Time or Daylight-
Saving Time—whichever is
applicable
*ECG Artifact Elimination
(Models 173 and 174 only)
off
*Heartbeat Coincidence
(Models 172, 173, and 174 only)
off
*Fetal Movement Detection
(if purchased and installed)
on
*Language
set according to shipping
destination
*Recorder Speed
United States: 3 cm/min
International: 1 cm/min
*Paper Scale
United States: 30–240 BPM
International: 50–210 BPM
*RS-232 Port 1
Communications Mode
HP
*RS-232 Port 1
Baud Rate
1200
*RS-232 Port 2
Communications Mode
ext. BP
*RS-232 Port 2
Baud Rate
600
*Remote Mark Annotation
on (
)
*HR Offset
(Models 172, 173, and 174 only)
on with 10-minute auto-revert
*UA Reference
10 relative units
FM
Summary of Contents for Corometrics 170 Series
Page 1: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Page 2: ......
Page 3: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Page 12: ......
Page 21: ...Revision D 170 Series Monitor 1 9 2003023 001 Safety Definitions of Symbols For your notes ...
Page 22: ...1 10 170 Series Monitor Revision D 2003023 001 Safety Definitions of Symbols ...
Page 42: ...3 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 58: ...4 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 78: ...6 6 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 106: ...10 4 170 Series Monitor Revision D 2003023 001 Supplies Accessories ...
Page 111: ......