Revision D
170 Series Monitor
5-11
2003023-001
Fetal Heart Rate Monitoring:
Fetal Heart Rate Alarms
FHR Low Alarm
The simplest example of a low FHR alarm occurs when the FHR value is
continuously
less than the threshold (low limit) for 30 seconds. When data
consistently violates the limit, the time-to-alarm is 30 seconds. See
Figure 5-6. Low FHR Alarm Example
Sample Clinical Exceptions
provides an example of FHR fluctuations above and below the high
alarm limit setting.
Figure 5-7. Fluctuations Near High Alarm Limit Example
Whether or not the pattern shown in
generates an alarm depends on what
percentage of the data violates the limit. The monitor evaluates the data on an on-
going basis; the methodology can be simplified as follows:
A FHR threshold alarm occurs if the FHR violates the alarm limit setting for
more time than it stays within the specified acceptable range.
The time-to-alarm increases as a greater percentage of data stays within the
specified acceptable range.
1
2
3
4
5
Time (minutes)
FHR (BPM)
low alarm limit
(threshold)
alarm activates
1
2
3
4
5
Time (minutes)
FHR (BPM)
high alarm limit
(threshold)
Summary of Contents for Corometrics 170 Series
Page 1: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Page 2: ......
Page 3: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Page 12: ......
Page 21: ...Revision D 170 Series Monitor 1 9 2003023 001 Safety Definitions of Symbols For your notes ...
Page 22: ...1 10 170 Series Monitor Revision D 2003023 001 Safety Definitions of Symbols ...
Page 42: ...3 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 58: ...4 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 78: ...6 6 170 Series Monitor Revision D 2003023 001 For your notes ...
Page 106: ...10 4 170 Series Monitor Revision D 2003023 001 Supplies Accessories ...
Page 111: ......