Factory Commands - Query Only
© 2010 Baker Hughes, a GE company – All rights reserved.
8000/8100/8200 Series Instruction Manual-English | 29
Serial Number = <Serial Number> Calibration Date= <Calibration
Date><CR>
Curve Fit = <X order> x <Y order><CR>
<CR>
Type <CR> to continue<CR> (Waits 20 seconds or for CR)
4.8.3 T - Analogue/Digital Converter Set-up
This direct mode only command queries the internal analogue to digital converter used to
measure the ambient temperature value.
Command:
n/a
Query:
<SPACE>T,?<CR>
Reply:
<Diode Calibration><CR>
Query:
<SPACE>*T,?<CR>
Reply:
Diode Cal = <Diode Calibration><CR>
4.8.4 V - Transducer Type and Settings
This direct mode only command queries the transducer type, pressure range and serial
number.
Command:
n/a
Query:
<SPACE>V,?<CR>
Reply:
<Type>,<Serial Number>,<Sensor Serial Number>,<Style>,<Pressure
Unit Number>,<Minimum Pressure>,<Maximum Pressure><CR>
Query:
<SPACE>*V,?<CR>
Reply:
Type = <Type><CR>
Table 4-5: Transducer Settings
Setting
Description
Type
5 Character text string.
Serial Number
‘AA/1/2’ format.
Sensor Serial Number
Serial number is a 7 digit integer.
Style
G for Gauge, A for Absolute.
Pressure Unit Number
0 to 24, see the U command.
Minimum Pressure
Real number in pressure units.
Maximum Pressure
Real number in pressure units.