PIN Protected Set-up Commands
© 2010 Baker Hughes, a GE company – All rights reserved.
8000/8100/8200 Series Instruction Manual-English | 27
Command:
<SPACE>*P, <Current PIN>,<New PIN><CR>
Reply:
n/a
Result:
The PIN of the transducer changes to new PIN.
Query:
<SPACE>P,?<CR>
Reply:
(Y if PIN set, N if PIN not set)<CR>
Query:
<SPACE>*P,?<CR>
Reply:
Pin Set = (Yes if PIN set, No if PIN not set)<CR>
4.7.6 S - Set Offset
This command adds an offset value to the output of the transducer. See Section 4.7.2 on
page 23 for more details.
Command:
<SPACE>S, <PIN>,<Pressure><CR>
Reply:
n/a
Result:
The transducer stores the value set with the S command in non-
volatile memory.
Query:
<SPACE>S,?<CR>
Reply:
<Off-set>,<Set-point><CR>
Query:
<SPACE>*S,?<CR>
Reply:
Offset = <Off-set><Units><CR>
Set At = <Set-point><Units><CR>
The transducer stores the value entered as <Set-point>. The H command also uses this set-
point value. To remove the set-point send:
Command:
<SPACE>S,<PIN>,X<CR>
Result:
Clears the set-point value.
where: <Current PIN> is the PIN of the transducer.
<New PIN>
is the replacement PIN.
INFORMATION
Proceed with caution when using this command.
Do not use the H and S commands together.