Chapter 2. RPS Pressure Measurement
6 | 8000/8100/8200 Series Instruction Manual-English
© 2010 Baker Hughes, a GE company – All rights reserved.
This equation is in the form:
The coefficients
K
i, j
,
X
and
Y
are printed on the calibration certificate supplied with the
pressure sensor. A sample set is shown in Table 2-1:
2.4 Stored Coefficients
The coefficients are also stored internally on a serial EEPROM. See Appendix B. for details on
data format and communication information.
where:
P
=
Applied pressure in mbar
K
i, j
= Calibration coefficient
x
=
Pressure signal in Hz
X
=
Pressure normalizing factor
y
=
Temperature signal in mV
Y
=
Temperature normalizing factor
Table 2-1: Sample Coefficients
Sample Coefficients
Sample Coefficients
Coefficient
Value
Coefficient
Value
K00
9.02
K31
2.975355E-14
K01
-8.654275E-02
K32
-1.591914E-16
K02
3.705644E-05
K33
-3.095734E-18
K03
-3.071498E-08
K34
0.00
K04
0.00
K40
4.689744E-15
K10
3.792730E-01
K41
-1.867269E-18
K11
4.884866E-06
K42
-2.591512E-20
K12
-8.219704E-09
K43
6.066456E-23
K13
-3.283229E-11
K44
0.00
K14
0.00
K50
-2.043712E-20
K20
9.252440E-06
K51
-4.652603E-21
K21
4.893925E-11
K52
2.124089E-23
K22
2.872573E-14
K53
3.812421E-25
K23
-1.617304E-15
K54
0.00
K24
0.00
X
2.04
K30
1.185548E-10
Y
5.02
P
K
i j
x X
–
i
y Y
–
j
j
0
=
4
i
0
=
5
=