Chapter 4. Programming Guide
30 | 8000/8100/8200 Series Instruction Manual-English
© 2010 Baker Hughes, a GE company – All rights reserved.
Serial Number = <Serial Number><CR>
Sensor SN = <Sensor Serial Number><CR>
Style = <Style><CR>
Minimum Pressure = <Minimum Pressure> <Pressure Units><CR>
Maximum Pressure = <Maximum Pressure> <Pressure Units><CR>
4.8.5 W - EEPROM Query
This direct mode only command queries the entire contents of the non-volatile memory.
Command:
n/a
Query:
<SPACE>W,?<CR>
Reply:
<Byte 0>,<Byte 1>,<Byte 2>,<Byte 3>,<Byte 4>, <Byte5>,<Byte 6>, … ,
<Byte 510>,<Byte 511><CR>
Query:
<SPACE>*W,?<CR>
Reply:
000=> <Byte 0>,<Byte 1>, … ,<Byte 15><CR>
010=> <Byte 16>,<Byte 17>, … ,<Byte 31><CR>
…
0F0=> <Byte 240>,<Byte 241>, … ,<Byte 255><CR>
<CR>
Send <CR> to continue<CR>
(Waits 20 seconds or for CR)
100=> <Byte 256>,<Byte 257>, ... ,<Byte 271><CR>
...
1F0=> <Byte 496>,<Byte 497>, … ,<Byte 511><CR>
Send <CR> to continue<CR>
(Waits 20 seconds or for CR)
4.9 Error Messages
Error messages are generated from the following sources:
4.9.1 Incorrect User Command
In a command string, the transducer accepts correct commands and rejects incorrect
commands. An incorrect command causes the transducer to send an error message related
to the command.
The form of the error message:
Reply:
ERROR nn