101 N. Alloy Dr.
Fenton, MI 48430
Ph (810)
714-5811
Fax (810) 714-5711
Research, Development and Manufacturing of Precision Measuring Systems
Form: CA 033
September 21,
2004
R:\Quality\Calibration Instructions\CA 033.doc Rev. A
Page 1 of 7
Configuration and Mastering Instruction for the
LMI 200-S or LMI 200-SB to the LMI 440 or ASI DataMyte 501
This process will outline:
Section
Pages
I.
Configuration of the LMI 440 or ASI DataMyte 501
2-4
II.
Mastering the LMI 200-S or 200-SB
5-6
III.
How to verify the mastering of the LMI 200-S or 200-SB
7
Required Equipment:
1
LMI 440 or ASI DataMyte 501
2
LMI 200-S or 200-SB transducer
3
LMI 210 master block
4
LMI 6009 4 pin – 4 pin cable
1
4
2
3