“Master” will be highlighted in screen header and “G4” is identified as “Gage Port”. If G4 is
not the Gage Port press the
►►
or
◄◄
until G4 appears.
7
Place the transducer into the top or “Lo” step of LMI 210 Master Block. Verify “Master Lo”
is highlighted on the collector, press <enter>.
8
Place the transducer into the bottom or “Hi” step of the LMI 210 Master Block. Verify
“Master Hi” is highlighted on the collector, press <enter>.
9
Position the transducer into the calibration block’s center or Master. Verify “Master Zero” is
highlighted on the collector, press <enter>.
Calibration/ Mastering for the LMI 200 Series is now complete.