Programming Site Data
2-17
September 2004
Setting Up a Totalizer
Output
1.
At
Output B
or
D
, select
OFF
or
TTLZR
and press
[ENTER]
.
2.
Do one of the following:
•
If you selected
OFF
, the meter returns to the
Global I/O
menu.
Go to
Configuring the Communications Port
on the next page.
•
For 1-channel meter, go to step 4.
•
For a 2-channel/path meter, go to step 3.
3.
At
Channel
, press [
X
] to select the desired channel and press
[ENTER]
.
4.
Press [
X
] to select the desired measurement parameter and press
[ENTER]
.
5.
Enter a value for the minimum pulse on-time (between 50
μ
sec
and 500,000
μ
sec) for the frequency of the totalizer pulses and
press
[ENTER]
. Press [
W
]
and [
X
] to move the cursor to the
desired location and press
[
S
]
and
[
T
]
to increase or decrease
the number. Then press
[ENTER]
.
Note:
A complete pulse consists of equal amounts of
ON
and
OFF
times. Choose a value that is compatible with the pulse
counter to be used.
6.
At
UNITS/PULSE
enter a value for the number of measurement
units represented by each pulse and press
[ENTER]
.