September 2004
2-10
Programming Site Data
KV Input Selection
1.
Press [
W
]
and [
X
] to enter a static
kinematic viscosity
or a table of
values and press
[ENTER]
.
2.
Do one of the following:
•
If you selected
Table
, enter the
Calibration Factor
by using the
arrow keys. Press [
W
]
and [
X
] to move the cursor to the
desired location and press
[
S
]
and
[
T
]
to increase or decrease
the number. Then press
[ENTER]
.
•
If you selected
STATC
, the AT868 will select and automatically
display the Kinematic Viscosity (see note below). If you want
to enter a different number use the [
W
]
and [
X
] to move the
cursor to the desired location and press
[
S
]
and
[
T
]
to
increase or decrease the number. Then press
[ENTER]
. Enter the
Calibration Factor
in a similar manner and press
[ENTER]
.
Note:
If you select
OTHER
as the fluid type, you must enter the
Kinematic Viscosity. The
AT868
will only approximate the
Kinematic Viscosity if you select one of the specific available
fluid options.
3.
Proceed to one of the following sections:
•
Clamp-on Transducers
- proceed to
Number of Traverses and
Transducer Spacing
below.
•
Wetted Transducers
- The meter returns to the
Channel
PROGRAM
window. To complete initial setup of the AT868, you
must enter data in the
GLOBL-SYSTM
submenu on page 2-12.
Number of Traverses and
Transducer Spacing
1.
Press [
W
]
and [
X
] to select the number of traverses and press
[ENTER]
.
2.
Enter the value for the transducer spacing by using the arrow keys.
Press [
W
]
and [
X
] to move the cursor to the desired location and
press
[
S
]
and
[
T
]
to increase or decrease the number. Then
press
[ENTER]
.