Checkout
2001989-351A
ApexPro™
7-11
6.
Read the current leakage indicated on DMM.
NOTE
If either reading is greater than the appropriate specification below, the
device under test fails. Contact GE Technical Support.
300 µA (0.3 volts on the DMM), and the device under test is powered from
100-120 V/50-60 Hz
300 µA (0.3 volts on the DMM), and the device under test is powered from a
centered-tapped 200-240 V/50-60 Hz, single phase circuit
500 µA (0.5 volts on the DMM), and the device under test is powered from a
non-center-tapped, 200-240 V/50-60 Hz, single-phase circuit
7.
Set the
GND
switch on the leakage tester to
CLOSED
.
8.
Read the current leakage indicated on DMM.
9.
Set the polarity switch to
RVS
.
10. Read the current leakage indicated on DMM.
NOTE
If the reading is greater than the specification below, and the device under
test is powered from 100-240 V/50-60 Hz, the device under test fails.
Contact GE Technical Support.
100 µA (0.1 volts on the DMM), and the device under test is powered from
100-240 V/50-60 Hz
11. Set the power switch of the device under test to OFF.
Test completion
1.
Disconnect the leakage tester from the power outlet.
2.
Disconnect all test equipment from the device.
3.
Disconnect the device power cord from the leakage tester.
Transmitter checkout
Checkout procedure
NOTE
Due to IPX7 rating, the ApexPro CH and T14 Transmitters are not field
repairable.
1.
Perform visual inspection.
See Visual inspection on page 4-4.
2.
Perform the communication tests.
See Communications tests on page 7-17.
3.
If an oximeter is being used with the transmitter, perform the Oximeter
operational tests for the applicable oximeter.
Summary of Contents for ApexPro TT
Page 9: ...2001989 351A ApexPro vii...
Page 10: ...viii ApexPro 2001989 351A...
Page 11: ...2001989 351A ApexPro 1 1 1 Introduction...
Page 20: ...1 10 ApexPro 2001989 351A Introduction...
Page 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview...
Page 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration...
Page 62: ...3 22 ApexPro 2001989 351A Installation and configuration...
Page 63: ...2001989 351A ApexPro 4 1 4 Maintenance...
Page 84: ...4 22 ApexPro 2001989 351A Maintenance...
Page 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting...
Page 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts...
Page 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module...
Page 139: ...2001989 351A ApexPro 7 1 7 Checkout...
Page 163: ...2001989 351A ApexPro A 1 A Technical specifications...
Page 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility...
Page 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility...
Page 191: ...2001989 351A ApexPro 1 1...