4-16
ApexPro™
2001989-351A
Maintenance
4.
Configure the ASCII setup.
a.
Select
File > Properties
.
b.
In the
Settings
tab, click
ASCII Setup...
c.
Verify the settings are the same as shown:
401A
d.
Click
OK
.
5.
Press
Enter
.
6.
When prompted, type the password
mms_aps
and press
Enter
.
7.
At the
@>
prompt, type
ss
and press
Enter
.
8.
Determine if patients are admitted to the receiver, use the following graphics for
reference:
If the column next to FQ is all -1 values, no beds are admitted on the receiver
subsystem tested. Proceed to step
Baud-Rate
19200
Data Bits
8
Parity
None
Stop Bit
1
Flow Control
Xon Xoff
Summary of Contents for ApexPro TT
Page 9: ...2001989 351A ApexPro vii...
Page 10: ...viii ApexPro 2001989 351A...
Page 11: ...2001989 351A ApexPro 1 1 1 Introduction...
Page 20: ...1 10 ApexPro 2001989 351A Introduction...
Page 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview...
Page 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration...
Page 62: ...3 22 ApexPro 2001989 351A Installation and configuration...
Page 63: ...2001989 351A ApexPro 4 1 4 Maintenance...
Page 84: ...4 22 ApexPro 2001989 351A Maintenance...
Page 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting...
Page 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts...
Page 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module...
Page 139: ...2001989 351A ApexPro 7 1 7 Checkout...
Page 163: ...2001989 351A ApexPro A 1 A Technical specifications...
Page 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility...
Page 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility...
Page 191: ...2001989 351A ApexPro 1 1...