7-22
ApexPro™
2001989-351A
Checkout
4.
Turn off the Accutracker.
5.
After 5 seconds (to prevent the unit from “locking up”), press and hold the
button and turn the Accutracker on.
6.
Press the
button twice; then press the
button to force the Accutracker to
activate the pump and valve circuitry. The unit goes through a diagnostic self-
test. If the unit passes the test, TEST PASSED displays briefly.
7.
When OFFICE SELF TEST displays, immediately press the
button to
discontinue the reading in progress. This step is necessary to identify that a
reading can be aborted, if necessary, during a test in case there is a problem (i.e.,
the cuff slips down the arm, an electrode becomes detached, etc.).
8.
Turn the Accutracker off to exit this mode.
Over-pressure release check
The Over-Pressure release occurs whenever a pressure of 285 mmHg (+/–20 mmHg)
is applied to the Accutracker pressure transducer. The Over-Pressure release forces
the pump to stop and valve #1 to open.
1.
Press and hold the
button and turn the Accutracker on.
NOTE
Make sure no external pressure is applied to the Accutracker when you turn
it on.
2.
Apply a pressure of 285 mmHg (+/–20 mmHg) to the Accutracker. Valve #1
immediately opens and allows the applied pressure to release.
3.
Once the Over-Pressure circuit is tripped, valve #1 and the pump are disabled
until “Analog Power” (+/–Vana) is removed from the circuit. Turn the
Accutracker off to reset the circuit.
Hardware time-out and system leak check
1.
Press and hold the
button and turn the Accutracker on.
NOTE
Make sure no external pressure is applied to the Accutracker when you turn
it on.
2.
Apply 200 mmHg pressure to the Accutracker pressure circuit and start a timer to
clock the Hardware Time-Out circuit.
3.
The Hardware Time-Out occurs in 3 minutes (+/–45 seconds). Valve #1 opens
and dumps pressure once the time-out is finished. As you monitor the Time-Out,
monitor the pressure (in mmHg) displayed on the Accutracker LCD. The
Accutracker should leak no more than 2 mmHg per minute.
Summary of Contents for ApexPro TT
Page 9: ...2001989 351A ApexPro vii...
Page 10: ...viii ApexPro 2001989 351A...
Page 11: ...2001989 351A ApexPro 1 1 1 Introduction...
Page 20: ...1 10 ApexPro 2001989 351A Introduction...
Page 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview...
Page 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration...
Page 62: ...3 22 ApexPro 2001989 351A Installation and configuration...
Page 63: ...2001989 351A ApexPro 4 1 4 Maintenance...
Page 84: ...4 22 ApexPro 2001989 351A Maintenance...
Page 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting...
Page 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts...
Page 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module...
Page 139: ...2001989 351A ApexPro 7 1 7 Checkout...
Page 163: ...2001989 351A ApexPro A 1 A Technical specifications...
Page 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility...
Page 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility...
Page 191: ...2001989 351A ApexPro 1 1...