This procedure verifies that the telemetry system transmits and receives the data
correctly from the Accutracker.
1.
Attach the Accutracker to the ApexPro Telemetry system.
NOTE
You must have ECG leads with a shorting cable or simulator attached to the
ApexPro Telemetry system for this test to work.
CAUTION
Refer to the ApexPro operator’s manual for proper operation
guidelines and cuff/microphone placement.
2.
Place blood pressure cuff on the arm.
3.
Turn the Accutracker on and press the
button.
4.
When the display appears on the system screen, verify that the Accutracker
display numbers match the display numbers.
5.
Turn Accutracker off.
Repair log
Unit Serial Number:
Institution Name:
Date
Maintenance/Repair
Technician
Summary of Contents for ApexPro TT
Page 9: ...2001989 351A ApexPro vii...
Page 10: ...viii ApexPro 2001989 351A...
Page 11: ...2001989 351A ApexPro 1 1 1 Introduction...
Page 20: ...1 10 ApexPro 2001989 351A Introduction...
Page 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview...
Page 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration...
Page 62: ...3 22 ApexPro 2001989 351A Installation and configuration...
Page 63: ...2001989 351A ApexPro 4 1 4 Maintenance...
Page 84: ...4 22 ApexPro 2001989 351A Maintenance...
Page 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting...
Page 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts...
Page 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module...
Page 139: ...2001989 351A ApexPro 7 1 7 Checkout...
Page 163: ...2001989 351A ApexPro A 1 A Technical specifications...
Page 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility...
Page 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility...
Page 191: ...2001989 351A ApexPro 1 1...