A
B
WARNING
Emergency stop. Pressing the EMERGENCY STOP will not shut
off mains power to the electrical cabinet.
Pressing the EMERGENCY STOP button (B) will result in the following:
•
Pump motors stop immediately.
•
The built-in computer and other components will remain powered.
•
No data will be lost.
•
The valves will shut down in a specific order, for details see
.
2 Safety instructions
2.3 Emergency procedures
ÄKTA ready XL Operating Instructions 29281616 AE
25