1.5 Introduction
The PC6-IDOS enables the user to carry out checks on
pressure transducers, transmitters, pressure switches, pressure
gauges and process indicators quickly and accurately. It is
intended for use as a portable, on-site calibrator, used in close
proximity with the equipment under test and without the need to
remove the unit from its parent system. This section describes
the connections to be made when checking switches,
transmitters, etc. It also lists menu options, selection procedures
and purpose of options available.
1.6 Connections
1.6.1 Pressure Input
The PC6-IDOS measures pressure
directly via the connector found underneath the case (i.e. the
connection between the PC6-IDOS and a hand held pressure
test system). Alternatively, connect the item under test to the
PC6-IDOS with the adaptor and seals provided.
1.6.2 EPM (External Pressure Module)
Different pressure ranges can be read by attaching an
External Pressure Module to the 5-way socket.
(See Fig.3 opposite)
Warning:
The PC6-IDOS must be switched off prior to
connecting or disconnecting the EPM .
1.6.3 Voltage / Current Input / Output
Voltage and Current
readings can be taken by connecting the outputs from the item
under test to the 'V' or 'mA' and 'GND' input sockets on the
connection panel using the connection leads provided.
Warning:
When measuring voltage or current, make sure
'GND' is connected (where applicable, see fig. 3a)
Summary of Contents for Druck PC6-IDOS
Page 9: ...DESCRIPTION 1 Page ...
Page 17: ...OPERATING INSTRUCTIONS 9 Page ...
Page 22: ...HOW TO SELECT FUNCTIONS MENU 1 14 Page ...
Page 28: ...SET MENU MENU 2 20 Page ...
Page 44: ...FLOWCHARTS Page 36 ...
Page 63: ...APPENDIX 1 External Pressure Module EPM 55 Page ...
Page 66: ...APPENDIX 2 Calibration Procedures 58 Page ...
Page 70: ......