To output voltage or current:
• Select ‘Source’ using the
key and press the
key.
Three options become available; ‘Actual’, ‘Convert’ and ‘P
V’ / mA.
• ‘Actual’ is selected by pressing the
key.
An example display below shows the Source Mode symbol on
the second line.
• Select ‘Converted’ using the
key and then pressing
the
key. To set the converted parameters refer to
section 2.2.1 on page 23.
• Select ‘P V’ /mA using the
key and then press the
key. To set the parameters for this Pressure to Volts /
milliamps conversion refer to section 2.2.3 on page 23.
An external pressure source is required.
• By pressing and holding the
key for more than two
seconds whilst in ‘Source Mode’ you will activate a ‘
’
function allowing the source value to be increased or
decreased, the longer you hold down the
or
key,
the larger the nudge increments will be (see Page 25 for
more information).
By pressing and holding the
key again for more than
two seconds, the nudge function will be cancelled.
Note:
Nudge
MENU
MENU
ENTER
ZERO
Page 16
Note: NOT Available in Ramp or
P-V / P-mA modes
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO