![GE HEALTHCARE Logiq Book XP series Basic Service Manual Download Page 94](http://html.mh-extra.com/html/ge-healthcare/logiq-book-xp-series/logiq-book-xp-series_basic-service-manual_571895094.webp)
GE H
EALTHCARE
D
IRECTION
5194296-100, R
EVISION
10
LOGIQ B
OOK
XP S
ERIES
B
ASIC
S
ERVICE
M
ANUAL
3-38
Section 3-9 - Loading the System Software
3-9-3
Loading the System Software with USB Memory Stick (For R2.2.2 or above)
NOTE:
Because any data may be lost or impacted during upgrade process, the preserved patient data
on the system should be backed up before application software upgrade.
1.) Disconnect the system from the network.
2.) Power off the system.
3.) Insert USB memory stick labeled “System & Application Software” to the system.
4.) Turn on the system. System will detect the USB memory stick automatically.
5.) Press any key to continue when below message display as shown in
Figure 3-36 on page 3-
38
.
Figure 3-36 System Software load message
Summary of Contents for Logiq Book XP series