4-6
Section 4-3 - General Procedure
4-3-4
Adjusting the Display Monitor
4-3-4-1
Brightness
To adjust the brightness:
Adjust the Brightness Slide pots, located on the right of side of the LCD, see
Figure 4-5 on page 4-6
.
Record the final brightness setting and leave this information with the system.
4-3-5
Lockout/Tagout (LOTO) requirements
Follow OSHA Lockout/Tagout requirements (USA) or local Lockout/Tagout requirements by ensuring
you are in total control of the AC power plug at all times during the service process.
To apply Lockout/Tagout:
1.) Plan and prepare for shutdown.
2.) Shutdown the equipment.
3.) Isolate the equipment.
4.) Apply Lockout/Tagout Devices.
5.) Remove battery.
6.) Control all stored and residual energy.
7.) Verify isolation.
All potentially hazardous stored or residual energy is relieved.
Figure 4-5 Brightness Slide pots
NOTICE
Energy Control and Power Lockout for LOGIQ Book XP Series
WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE
GREATER THAN 30 VOLTS:
1. TURN OFF THE SCANNER.
2. UNPLUG THE SYSTEM.
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE NO
TEST POINTS TO VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF BUTTON
WILL TURN OFF.
5. REMOVE THE SYSTEM BATTERY.
Signed
Date
TAG
&
LOCKOUT