Chapter 7 Diagnostics/Troubleshooting
7-23
7-6-2-4
DCNN
These programs are provided for testing the DCNN board.
The PC sends the command to the SH4 in the MST, which brings DCNN into test then gets test report
from DCNN via DUSC Bus and sends it back to the PC.
Table 7-6
FRU Test, DCNN Diagnostics Menu
Menu
Descriptions
Error Code
Overall Test
Performs all of the DCNN board tests listed below.
Assy Revision
Test
N/A
40001