Chapter 7 Diagnostics/Troubleshooting
7-21
7-6-2-2
DFP
These programs are provided for testing the DFP board.
Table 7-4
FRU Test, DFP Diagnostics Menu
Menu
Descriptions
Error Code
Overall Test
Performs all of the DFP board tests listed below.
Assy Revision
Test
20001
DUSC I/F Test
Check if the SH4 in the MST can access the DFP board via DUSC bus.
20002
Local Bus Test
Check if SH4 can read and write register of DFP board via DUSC bus.
20003
WRAM Test
The SH4 in MST reads and writes the register of WRAM in DFP via DUSC bus.
20004
OQCARD Test
The SH4 in MST reads and writes the OQCARD registers in DFP via bus.
20005