Chapter 7 Diagnostics/Troubleshooting
7-15
7-4-5
Update FPGA
15.)Choose Utilities in Global Service Interface.
16.)Click “LOGIQ Book XP FPGA Update” to active Fpga update tool.
17.)Click “Update” to Fpga Update tool.
18.)Restart system after completing Fpga upgrade.
7-4-6
Control Frame
Contains the user interface elements used for:
•
Diagnostic control, and
•
Operator feedback
7-4-6-1
Button
This button has two modes each with appropriate text:
•
Execute - to start the diagnostic, and
•
Abort - to stop a diagnostic
The button can also be disabled.
7-4-6-2
Loop Count
This is an editable text field that will only accept numeric values with 4 digits or less. When the button
is configured as an “execute” button and pressed, the loop count field will be queried to determine the
number of times to execute the diagnostic.
Figure 7-15
Figure 7-16 Reset Database