  Visually check if the product or the medical gas cylinder is contaminated; if needed, 

for the regulator, use the cleaning procedure detailed in this section (if required for 
the cylinder, refer to the gas cylinder manufacturer cleaning recommendation). 

  Check if the product service is due or that the total life time of the product and the 

gas cylinder has not been exceeded, (refer to GCE or owner’s date coding system). 
If service or life time has been exceeded, remove the product (or the gas cylinder) 
from service & suitably identify its status.

 

  Ensure that the product inlet stem is compatible with the medical cylinder valve (gas/ 

thread type). 

  Check  the presence & the integrity of inlet stem seals / correct size of seal.

 

 

 

 

Remove  caps  from  inlet  and/or  flow  outlet.  Keep  caps  in  a  safe  place  for  reuse 
during transport or storage. 

 

 

The product is dedicated only for use with the gas specified on its labelling.  Never 
try to use for another gas. 

  Secure the gas cylinder stand. 

 

Screw connection (bull nose or nut type) 

  Manually screw the bull nose or the nut onto the cylinder valve connector. 

  Turn the regulator into the correct position for use and tighten the nut by hand - do 

not use tools. 

 
Pin index connection 

  Position  the  pin-index  over  the  cylinder  valve  with  the  pin(s)  on  the  pressure 

regulator pointing towards the cylinder valve connector holes on the cylinder valve. 

  Press the regulator inlet connection pins into the cylinder valve connector holes - do 

not use force, otherwise the pins or holes may be damaged.  

  Tighten  the  screw  on  the  regulator  onto  the  cylinder  valve  connector  via  the  T-bar 

handle. Do not use tools. 

 

  Position the equipment so that the regulator user outlets point away from personnel 

or patient. 

 

 

Fitting the regulator with too high a torque to the cylinder valve may result in 
damage. 
 

 

During fitting to the cylinder valve, do not apply torque/load to any other parts 
of the product.  

  For  regulators  fitted  with  a  flow-metering  device,  set  the  flow  control  knob  on  the 

“ZERO” position - Ensure the flow control knob engages correctly. 

  Open the cylinder valve slowly by turning the hand wheel in anticlockwise direction 

approx 1 to 1½ turns. 

 

 

Sudden  opening  of  the  cylinder  valve  could  result  in  a  danger  of  fire  or 
explosion  arising  from  oxygen  pressure  shocks.  Insufficient  opening  of  the 
cylinder valve could reduce actual flow delivered. 

 

Summary of Contents for mediline MEDIREG II

Page 1: ...Regulators IFU GB FR DE NL Rev 03 Page 1 44 12 02 2010 MEDIREG II MEDISELECT II MEDICAL HIGH PRESSURE REGULATORS Document Nr 735 800 000 092 Date of issue 12 02 2010 Revision Nr 3 GB FR DE NL...

Page 2: ...thing helium carbon dioxide xenon specified mixtures of the gases listed air or nitrogen to power surgical tools Operations Transport Storage Keep the product and its associated equipments away from h...

Page 3: ...alve due to that the gas is going from high pressure to low pressure Joule Thompson effect Ensure that all equipment that is connected to the valve by at least a 2 metre hose According to Medical Devi...

Page 4: ...the pressure outlet can be used Pressure outlet I is fitted with a gas specific medical quick connector also called quick coupler The user can connect another piece of equipment to this outlet with a...

Page 5: ...nto the cylinder valve connector Turn the regulator into the correct position for use and tighten the nut by hand do not use tools Pin index connection Position the pin index over the cylinder valve w...

Page 6: ...sition Do not use excessive force Reset the flow control knob to on the ZERO position once the gas flow stops and the regulator is vented For regulators fitted with a pressure outlet ensure it is func...

Page 7: ...position Ensure that the accessory is connected to the flow outlet Slowly open the cylinder valve by turning the hand wheel in anticlockwise direction approx 1 to 1 turns Sudden opening of the cylind...

Page 8: ...hand wheel in a clockwise direction to the stop position Do not use excessive force Reset the flow control knob on the ZERO position when the gas venting has ceased valid for version with flow meterin...

Page 9: ...ears At the end of the product s life time the product must be withdrawn from service The owner shall put in place a relevant procedure to ensure the product cannot be used again 2 Repairs activities...

Page 10: ...pressure range P2 Outlet pressure P4 Max outlet pressure closing pressure Q Outlet flow Service or disposal date The serial number indicates the year the product has to undergo the overall maintenance...

Page 11: ...ent that it is modified serviced or repaired by personnel not employed or authorised by GCE or if the apparatus is used in a manner not conforming to its intended use GCE cannot be held responsible fo...