Regulators IFU GB,FR,DE,NL Rev 03 

Page 8/44           

 12/02/2010 

 

  Ensure that the flow control knob is on the ZERO position (if any). 

  Ensure the accessory IS NOT connected to the pressure outlet. 

  Slowly open the cylinder valve by turning the hand wheel in anticlockwise direction 

approx 1 to 1½ turns. 

 

 

Sudden  opening  of  the  cylinder  valve  could  result  in  a  danger  of  fire  or 
explosion  arising  from  oxygen  pressure  shocks.  Insufficient  opening  of  the 
cylinder valve could reduce actual flow delivered. 

 

  Connect the accessory to the pressure outlet. 

 

After completion of the therapy 

  Turn off the cylinder valve by turning the hand wheel in a clockwise direction to the 

stop position. Do not use excessive force. 

  Vent gas pressure from downstream equipment.

 

 

Disconnect the male QC probe from the pressure outlet. 

) !

  Turn off the cylinder valve by turning the hand wheel in a clockwise direction to the 

stop position. Do not use excessive force. 

  Reset the flow control knob on the “ZERO” position – when the gas venting has ceased  

(valid for version with flow-metering device only). 

  Ensure the pressure gauge does not show any residual pressure. 

  Remove connections from user outlets. 

  Refit pressure outlet and flow outlet protection caps. Before refitting the caps, ensure 

they are clean.

 

. $

Remove general contamination with a soft cloth damped in oil free oxygen compatible 
soapy water & rinse with clean water.  
Disinfection can be carried out with an alcohol-based solution (spray or wipes). 
If other cleaning solutions are used, check that they are not abrasive and that they are 
compatible the product materials (including labels) and gas.  
 

 

Do not use cleaning solutions containing ammonia! 

 

 

Do not immerse in water or any liquid. 

 

 

Do not expose to high temperature (such as autoclave).

 

/ 0

/

(

1

 

Form of nine digit serial number stamped on the product is following: 
YY MM XXXXX 
YY: year of production 
MM: month of production 

Summary of Contents for mediline MEDIREG II

Page 1: ...Regulators IFU GB FR DE NL Rev 03 Page 1 44 12 02 2010 MEDIREG II MEDISELECT II MEDICAL HIGH PRESSURE REGULATORS Document Nr 735 800 000 092 Date of issue 12 02 2010 Revision Nr 3 GB FR DE NL...

Page 2: ...thing helium carbon dioxide xenon specified mixtures of the gases listed air or nitrogen to power surgical tools Operations Transport Storage Keep the product and its associated equipments away from h...

Page 3: ...alve due to that the gas is going from high pressure to low pressure Joule Thompson effect Ensure that all equipment that is connected to the valve by at least a 2 metre hose According to Medical Devi...

Page 4: ...the pressure outlet can be used Pressure outlet I is fitted with a gas specific medical quick connector also called quick coupler The user can connect another piece of equipment to this outlet with a...

Page 5: ...nto the cylinder valve connector Turn the regulator into the correct position for use and tighten the nut by hand do not use tools Pin index connection Position the pin index over the cylinder valve w...

Page 6: ...sition Do not use excessive force Reset the flow control knob to on the ZERO position once the gas flow stops and the regulator is vented For regulators fitted with a pressure outlet ensure it is func...

Page 7: ...position Ensure that the accessory is connected to the flow outlet Slowly open the cylinder valve by turning the hand wheel in anticlockwise direction approx 1 to 1 turns Sudden opening of the cylind...

Page 8: ...hand wheel in a clockwise direction to the stop position Do not use excessive force Reset the flow control knob on the ZERO position when the gas venting has ceased valid for version with flow meterin...

Page 9: ...ears At the end of the product s life time the product must be withdrawn from service The owner shall put in place a relevant procedure to ensure the product cannot be used again 2 Repairs activities...

Page 10: ...pressure range P2 Outlet pressure P4 Max outlet pressure closing pressure Q Outlet flow Service or disposal date The serial number indicates the year the product has to undergo the overall maintenance...

Page 11: ...ent that it is modified serviced or repaired by personnel not employed or authorised by GCE or if the apparatus is used in a manner not conforming to its intended use GCE cannot be held responsible fo...

Reviews: