Pressure outlet II  

  Ensure the counterpart is compatible with the pressure outlet features. 

  Screw the counterpart. 

  Check the counterpart is fully screwed. 

 

 

When is pressure outlet used by medical product with high flow consumption 
(for  example  lung  ventilator  with  request  of  source  flow  100  l/min  at  minimal 
pressure 2,8 bar) check the required capacity of source device with regulator 
pressure  outlet  performance  listed  in  appendix  1.  To  obtain  enough 
performance  of  the  regulator  is  recomended  replace  cylinder  when  gauge 
reach the red area. 

 

When  connecting  any  accessory  to  the  flow  outlet  make  sure  that  it  is  not 
connected to the patient before operating the product. 

 

  Ensure the hose/humidifier is compatible with the flow outlet feature.

 

  Push the hose onto the regulator flow outlet/outlet for humidifier. 

  Ensure the hose/humidifier is well engaged. 

  Ensure that the flow control knob is on the ZERO position. 

  Ensure that the accessory is connected to the flow outlet. 

  Slowly open the cylinder valve by turning the hand wheel in anticlockwise direction 

approx 1 to 1½ turns. 

 

 

Sudden  opening  of  the  cylinder  valve  could  result  in  a  danger  of  fire  or 
explosion  arising  from  oxygen  pressure  shocks.  Insufficient  opening  of  the 
cylinder valve could reduce actual flow delivered. 

 

  Set the flow control knob on the required one of the available flow rates. 

 

 

Always  ensure  that  the  flow  control  knob  has  engaged  and  not  placed 
between  two  settings  otherwise  the  flow  selector  will  not  deliver  the  correct 
flow of medical gas. 

 

 

Do not try to apply an excessive torque on the flow control knob when it stops 
on the maximum flow position or in zero position. 

 

 

The  oxygen  flow  rate  must  be  prescribed  and  administered  by  a  clinically 
trained user. 

 

After completion of the therapy 

  Turn off the cylinder valve by turning the hand wheel in a clockwise direction to stop 

position. Do not use excessive force. 

  Vent gas pressure from downstream equipment. 

  Reset flow control knob on the ZERO position when gas venting has ceased. 

  Disconnect the tube/humidifier from the flow outlet.  

Summary of Contents for mediline MEDIREG II

Page 1: ...Regulators IFU GB FR DE NL Rev 03 Page 1 44 12 02 2010 MEDIREG II MEDISELECT II MEDICAL HIGH PRESSURE REGULATORS Document Nr 735 800 000 092 Date of issue 12 02 2010 Revision Nr 3 GB FR DE NL...

Page 2: ...thing helium carbon dioxide xenon specified mixtures of the gases listed air or nitrogen to power surgical tools Operations Transport Storage Keep the product and its associated equipments away from h...

Page 3: ...alve due to that the gas is going from high pressure to low pressure Joule Thompson effect Ensure that all equipment that is connected to the valve by at least a 2 metre hose According to Medical Devi...

Page 4: ...the pressure outlet can be used Pressure outlet I is fitted with a gas specific medical quick connector also called quick coupler The user can connect another piece of equipment to this outlet with a...

Page 5: ...nto the cylinder valve connector Turn the regulator into the correct position for use and tighten the nut by hand do not use tools Pin index connection Position the pin index over the cylinder valve w...

Page 6: ...sition Do not use excessive force Reset the flow control knob to on the ZERO position once the gas flow stops and the regulator is vented For regulators fitted with a pressure outlet ensure it is func...

Page 7: ...position Ensure that the accessory is connected to the flow outlet Slowly open the cylinder valve by turning the hand wheel in anticlockwise direction approx 1 to 1 turns Sudden opening of the cylind...

Page 8: ...hand wheel in a clockwise direction to the stop position Do not use excessive force Reset the flow control knob on the ZERO position when the gas venting has ceased valid for version with flow meterin...

Page 9: ...ears At the end of the product s life time the product must be withdrawn from service The owner shall put in place a relevant procedure to ensure the product cannot be used again 2 Repairs activities...

Page 10: ...pressure range P2 Outlet pressure P4 Max outlet pressure closing pressure Q Outlet flow Service or disposal date The serial number indicates the year the product has to undergo the overall maintenance...

Page 11: ...ent that it is modified serviced or repaired by personnel not employed or authorised by GCE or if the apparatus is used in a manner not conforming to its intended use GCE cannot be held responsible fo...

Reviews: