In case of storage at a temperature below -20 °C, d o not operate the regulator
until it has been allowed to increase its temperature to a minimum of -20 °C
•
At regulators intended for use with mixture of medical gases O
2
+N
2
O is lower
operation temperature limit +5°C. During normal use the flow outlet and
pressure outlet will sometimes have a frosty appearance. This is a normal
physical reaction in the valve, due to that the gas is going from high pressure
to low pressure (Joule Thompson effect). Ensure that all equipment that is
connected to the valve by at least a 2 metre hose.
According to Medical Devices Directive 93/42/EEC the product owner must ensure that
all personnel handling the product are provided with the operating instructions &
performance data and are fully trained to carry out that operation. Trainees need to be
supervised by an experienced person.