gbo Medizintechnik AG
Version 1.4
5
Electrodes
The standard accessories for HiToP
Metabol include rubber electrodes of 120 x
80 mm.
5.1 Electrode application
The positioning of the patient is very important. The therapy shall always be
carried out in a comfortable and relaxed position. The joints shall be located in an
angular position, so that they are placed in a mid-position, from which both the
flexor as well as the extensor muscles can be stimulated.
In general, the preparation and application of the electrodes is carried out as
follows:
1. Connect the electrodes to the patient cable.
2. Connect the patient cable to the distribution cable.
3. Plug in the patient plug of the distribution cable to the socket of the unit.
4. Wet the conductive rubber electrodes with Aloe contact spray to avoid current
sensations on the skin.
5. If you want to use the macroporous intermediate layers or fleece pockets, they
must be thoroughly washed under running water.
6. Apply the electrodes to the patient.
7. Apply the holding strap in a way that the electrode fits completely.
Warning!
Do not apply the electrodes on skin injuries. Even minor
abrasions can cause a burning to the patient. Herewith the
intensity of the current can be erroneously judged by the patient.
If this cannot be avoided, apply zinc ointment or vaseline to
cover the affected parts of the skin.
The electrodes must adhere completely to the skin in order to
avoid excessive local current density. Otherwise there can occur
erythemas, and sometimes even burns.