In accordance with the EMC regulations for medical products we are obliged by law to
provide the following information.

Guidance and manufacturer’s declaration — electromagnetic emissions

The equipment is intended for use in the electromagnetic environment specified below. The customer or the
user of the equipment should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions,
CISPR 11

Group 1

The equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.

RF emissions,
CISPR 11

Class B

The equipment is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.

Harmonic emissions,
IEC 61000-3-2 (*)

Class A

Voltage fluctuation/flicker
emissions, IEC 61000-3-3 (*)

Complies

(*) Note: For devices with a power consumption between 75 W and 1000 W only.

Guidance and manufacturer’s declaration — electromagnetic immunity

The equipment is intended for use in the electromagnetic environment specified below. The customer or the
user of the equipment should assure that it is used in such an environment.

Immunity test

IEC 60601- test level

Compliance level

Electromagnetic environment –

guidance

Electrostatic discharge (ESD),
IEC61000-4-2

6 kV contact

8 kV air

6 kV contact

8 kV air

Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.

Electrical fast transient/burst,
IEC 61000-4-4

2 kV for power supply

lines

1 kV for input/output

lines

2 kV for power supply

lines

1 kV for input/output

lines

Mains power quality should be that of a
typical commercial or hospital
environment.

Surge,
IEC 61000-4-5

1 kV differential mode

2 kV common mode

1 kV differential mode

2 kV common mode

Mains power quality should be that of a
typical commercial or hospital
environment.

Voltage dips, short
interruptions and voltage
variations on power supply
input lines,
IEC 61000-4-11

<5% U

for ½ cycle
(>95% dip)

40% U

for 5 cycles
60% dip)

70% U

for 25 cycles
30% dip)

<95% U

for 5 s
(>5% dip)

<5% U

for ½ cycle
(>95% dip)

40% U

for 5 cycles
60% dip)

70% U

for 25 cycles
30% dip)

<95% U

for 5 s
(>5% dip)

Mains power quality should be that of a
typical commercial or hospital
environment.

If the user of the equipment requires
continued operation during power mains
interruptions, it is recommended that the
equipment be powered by an
uninterruptible power supply or a
battery.

Power frequency (50/60 Hz)
magnetic field,
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.

Note: U

is the a.c. mains voltage prior to application of the test level.

Summary of Contents for HiToP Metabol

Page 1: ...High Tone Power Therapy Device HiToP Metabol User Manual...

Page 2: ...missions All rights reserved No part of this manual may be reproduced in any form or by any means electronic mechanical or otherwise without the prior written permission of the gbo Medizintechnik AG V...

Page 3: ...ol 3 gbo Medizintechnik AG Version 1 4 Warnings and safety precautions Warning Warnings which have to be observed by all means Caution Observe the instructions for use Note Information which facilitat...

Page 4: ...and installation 10 3 2 Connection and switch on 10 3 2 1 Voltage change over and fuses 11 3 3 Basic settings 12 3 3 1 Treatment time 12 3 3 2 Language 12 3 4 Important notes and safety precautions 13...

Page 5: ...of 4096 32768 Hz The therapy frequency is scanned with a defined frequency This method is called SimulFAM for Simultanous Frequency Amplitude Modulation In the following the basic method is divided i...

Page 6: ...l panel HiToP Int SimulFAM X Figure 1 Control panel 2 1 1 Ergonomic control panel The control panel possesses a display the white Softkey two function keys and the Intensity regulator These elements a...

Page 7: ...iew 2 1 4 Function keys HiToP Metabol is equipped with two black Function keys These are easy to feel and also have a pressure point With them you can scroll through the display indications so that yo...

Page 8: ...en the other electrode on the black side with Aloe Vera spray Then push the electrode down between leg and strap The black side of the electrode must point towards your skin Now fix the electrodes wit...

Page 9: ...bly several turns clockwise The therapy has started During treatment you can change the intensity While changing the intensity you will always notice a pulsation The display shows the treatment time r...

Page 10: ...portant notes and safety precautions Warning The unit is not designed to be used at non explosion proof places If it is used in dangerous areas of anesthesia departments the possibility of an explosio...

Page 11: ...iver the box can be pulled out of the receiver by the small slot Warning The mains socket is sealed with a scotch tape This is to remind to check prior to switch on whether the voltage set corresponds...

Page 12: ...ngs with the arrow keys After selection confirm your choice with the white softkey and you can modify the values 3 3 1 Treatment time The treatment time can be set in steps of 5 minutes between 30 and...

Page 13: ...he patient cable is in direct range of high frequency short wave or micro wave therapeutic devices a damage to the device or an injury of the patient cannot be excluded Please keep a distance of at le...

Page 14: ...treatment decrease the intensity to 0 by turning the intensity knob counterclockwise 4 2 Treatment details 1 Switch on the unit 2 You see the start display 3 In the display the following message appe...

Page 15: ...eps of seconds When the treatment timer reaches zero the unit stops the treatment and the intensity is set to zero The values for voltage U and current I can be used as an indicator for the following...

Page 16: ...ncy of treatment The duration of treatment is pre set to 60 minutes In the basic settings the treatment time can be set from 30 to 60 minutes Treatment of osteoarthrosis can be executed three times a...

Page 17: ...Plug in the patient plug of the distribution cable to the socket of the unit 4 Wet the conductive rubber electrodes with Aloe contact spray to avoid current sensations on the skin 5 If you want to use...

Page 18: ...arried out on the basis of this directive Here the operator regulation has to be especially observed Irrespective of the legal rules or beyond the scope of the Medical Device Directive it is recommend...

Page 19: ...me warm water Wash off the rest of the aloe vera contact spray Dry the electrodes with a cleaning tissue or let them air dry Note The conductivity of the electrodes is optimized by using a big part of...

Page 20: ...accordance to IEC 601 Protection against ingress of water IP X1 Dimensions 6 cm x 23 cm x 32 cm H x D x W Weight max 3 7 kg without accessories Color white RAL 9002 and grey RAL 7016 Display LCD back...

Page 21: ...on HiToP Metabol Mains cable Registration cards 017 0 0016 Distribution cable I red 150 cm 017 0 0030 Set of patient cables red 6 pcs 100 cm each 017 0 0034 Rubber electrodes 115 x 175 2 St ck 017 0 0...

Page 22: ...tient plug removed This error message appears when the connector is unplugged during therapy or the connector has a loose contact Error 3 current monitoring This error message appears when contact pro...

Page 23: ...by the manufacturer is 10 years The HiToP 191 is an electronic device For its disposal the respective regulations for electronic devices must be observed Incidentals have to be disposed with residual...

Page 24: ...ity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete...

Page 25: ...MHz to 800 MHz d 2 3 P for 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation dist...

Page 26: ...Connection and switch on 11 Disinfection 19 Display 7 F Function keys 8 Fuses 12 G Graphic display 7 H High Tone Therapy 6 I Intensity regulator 8 N note 15 S Safety controls 19 Softkey 8 Supply chan...

Reviews: