Duodynator
41
gbo Medizintechnik AG
Version 1.2
16
Annex I List of low frequency currents
Current form
Graphic
Pulse
duration in
ms
Pause
duration
in ms
DC
portion
in%
Frequency in
Hz
Maximum
Voltage Vpeak
Maximum
Current in
mApeak
Galvanic current
-
- 100
0
100
50
Ultra
Stimulation
2.0
5.0 28.6
143
100
70
Faradic current
1.0
19.0 2.5
50
100
100
DF
Diphase fixed
-
- 63
100
100
70
CP
Courtes périodes
-
- 47.25
50/100
100
70
MF
Monophasé
fixed
10
10 31.5
50
100
70
LP
Longues
périodes
-
- 46
50/100
100
70
MM
Monophasé
modules
-
- 20
100
100
70
RS
Rhythme
syncope
-
- 15.25
50
100
70
Impulse
galvanization
IG 30
30.0
50.0 18
12.5
100
70
Impulse
galvanization
IG 50
50.0
70.0 0.3
8.3
100
70
Frequency
modulation
FM / 7 - 14 Hz
0.5
70.9 to
142.4
0.4
7-14
100
100
Surged current
aS
-
- 32
0.2
100
70
Surged current
aSL
-
- 45
00:09
100
70
Tens
0.2 4.8 to 666 <3.5
1.5 to 200
100
100
Tens burst
1
4-666 <10
1.5 to 200
100
100
E100
100,0
500,0 7,14
1,67
100
100
E200
200,0
1000,0 7,14
0,83
100
100
!!
Note!
All low frequency currents except Galvanic current are also available
as metal- acceptable biphasic variant.