Troubleshouting
Each problem that occurs during operation of the unit will be shown in a message
window on top of the display and also signalized by an acoustic tone. Most of the
problems can be solved by the instructions displayed.
In general:
1.
The malfunction is shown on the display.
2.
An acoustic signal is heard.
3.
Follow the instructions on the display.
Suggestions:
Turn off and turn on the unit.
A full selftest of all device functions is performed and the unit is reinitialized. If
the unit does not come up to normal state, please call your local service.
13.1
Further errors
symptom
cause and action
Device cannot be switched on,
no display shown.
Please check the main plugs and sockets. If
necessary; contact your service agents or
the manufacturer.
No acoustic signal is heard.
(End of therapy…)
Please check the settings of the acoustic
tones in the menu. The volume must be
greater than 0.
Please contact your service agent or the manufacturer if the problems cannot be solved
by the measures mentioned above.
!!
Note!
Please note that the unit must be placed on a plane horizontal surface.
The device should neither be placed in front of radiators nor should it be
covered with pillows or blankets while in operation. Do not cover the
ventilation slots on the bottom of the unit either.