4.3
Implementation of the treatment
4.3.1
Treatment with interference currents
1.
Turn on the power switch on the back of the unit.
2.
The unit performs an automatic self-test of all functions.
3.
Error free self-test is completed by displaying the Home screen.
4.
Connect the accessory for interference currents to the socket on the right side, if
necessary, connect the cable of the
Neoserv
.
5.
Apply the electrodes to the patient (see section
.
6.
Press the button interference =
.
The last selected therapy is automatically
loaded.
7.
If necessary, change the therapy as described in chapter 4.2 "Selection of
therapies"
8.
If necessary, change the parameters and the duration for therapy.
9.
Turn the intensity up slowly until the interference current is clearly perceptible,
but not loosen muscular contraction duration from.
10.
At the end of treatment, the Therapy-end signal sounds and the intensity
automatically goes to "0.0".
4.3.2
Therapy with single channel currents
1.
Turn on the power switch on the back of the unit.
2.
The unit performs an automatic self-test of all functions.
3.
Error free self-test is completed by displaying the Home screen.
4.
Connect the accessory for single-circuit currents in the socket on the right side, if
necessary, connect the cable to the
Neoserv
.
5.
Apply the electrodes to the patient (see chapter).
6.
Press the button single channel =
.
The last selected therapy is
automatically loaded.
7.
Select a channel button channel I or II. The selected channel is connected to the
detail window.
8.
If necessary, change the therapy as described in chapter 4.2 "Selection of
therapies"
9.
If necessary, change the parameters and the duration for therapy.
10.
Turn on the intensity knob until the desired stimulus success occurs.
11.
At the end of treatment, the Therapy-end signal sounds and the intensity
automatically goes to "0.0".