Suction flyout
When using the
Duodynator
together with the
Neoserv
the suction operation will be
activated after the
Neoserv
is switched on. Pressing the suction button, a flyout
folded up in which you can set the parameters for the suction.
The suction, which means the negative pressure in the suction electrodes, can be
adjusted from -40 to -300 mbar.
The suction massage can be adjusted between 0 and 20 pulses / minute.
!
Note!
If the suction is to be used, the
Neoserv
must be switched on with its
power switch on the back of the unit before.
If the
Duodynator
is switched off using the power switch, the
Neoserv
sounds a warning signal which indicates that the
Neoserv
is
still on and must be switched off.
!
Note!
If the
Neoserv
is switched on and connected to the current
stimulation output of Duodynator, suction will automatically
switched on when selecting a current from the Favorite list.
Suction strength
suction massage
Suction on-off
switch
Suction
Button