Duodynator
21
gbo Medizintechnik AG
Version 1.2
4.4
Combination therapy with ultrasound
4.4.1
General information
The
Duodynator
may only be operated with the
Sonostat
of gbo Medizintechnik
AG. When operating the
Duodynator
in combination with other devices gbo
Medizintechnik AG assumes no product liability and warranty.
!!
Note!
If a different ultrasound device is connected, the combination have to be
approved according EN 60601-1 as a system.
The necessary connection between the two devices is the input jack on the back side
of the Sonostat. The device's internal flow path is activated for the selected
ultrasound probe.
Warning!
Observe the operating instructions of the ultrasonic therapy
device (Sonostat)!
The ultrasound therapy must not be operated with the current form
of electroplating due to the acidification of the contact gel.
A current density of 2 mA/cm
2
should not be exceeded.
Disable the current monitoring (Chapter 3.4 Settings) so that a
smooth course of therapy is possible.
!!
Note!
Combination therapy should be performed only with biphasic
currents.
We recommend medium frequency currents for the combination
therapy.
Use a conductive electrode gel to ensure a flow of current.