Duodynator
25
gbo Medizintechnik AG
Version 1.2
5.3
Suction electrodes
Instead of the above-mentioned electrodes, suction electrodes can be used for faster
application, in connection with the suction massage device. Single-pole suction
electrodes are applicable for all therapy currents. For this please plug in the
connection cable of the
Neoserv
to the socket of the current stimulation device
Duodynator.
Suction electrode
Surface
Maximum
current
Electrode plate
60 mm with sponge
60 mm
28 cm2
56 mA
Electrode plate
90 mm with sponge
90 mm
63.5 cm2
127 mA
Table 2: Maximum currents (rms) for suction electrodes