Gastops Ltd.
C009423, Rev. 4
page 51
All use of this publication and ChipCHECK by user shall be subject to the restrictions and conditions as set out in this publication.
The unit itself weighs approximately 40 lbs (18 kg). The Transport Case that was provided with the
instrument must be used any time the instrument is moved positions or locations. Where possible, two
people shall transport the unit and the instrument shall be kept in its vertical orientation at all times.
The following procedure must be followed when transporting the instrument:
1.
Complete an instrument shutdown in accordance with Section 4.7.
2.
Unplug the AC power cable from both the unit and the wall power socket.
3.
Open the provided Transport Case by undoing all latches and removing the top cover.
4.
Remove the instrument’s accessory foam insert from the Transport Case.
5.
Carefully lift the instrument and place it into the Transport Case, within its dedicated custom
foam cut out. Note that there is only one proper orientation for the unit to be placed, otherwise
the Transport Case will not close properly.
6.
Replace the instrument’s accessory foam insert, and place all of the instrument’s accessories
and power cable into their respective custom foam cut outs.
7.
Replace the Transport Case cover and re-fasten all latches.
8.
Lock or fasten each latch (where possible).
9.
Where possible, two people shall move the unit (in its transport case) to its new location. The
Transport Case is also equipped with a retractable handle and wheels for single person
transport.
NOTE:
Relocation within the same room is acceptable to be completed without using the transport
case; however, the transport case must be used otherwise. Prior to shipment/relocation, ensure
that the lid is fully and firmly attached to the
main body of the case with all latches fastened
or locked. All parts, components and accessories required to operate the unit are must be
included in the transport case.
NOTE:
The Sample Tray must be removed from the unit and placed in its proper location within the
foam insert prior to transport.