The unit itself weighs approximately 40 lbs (18  kg).  The Transport Case that was provided with the 
instrument must be used any time the instrument is moved positions or locations.  Where possible, two 
people shall transport the unit and the instrument shall be kept in its vertical orientation at all times.  
The following procedure must be followed when transporting the instrument: 

1.

 

Complete an instrument shutdown in accordance with Section 4.7. 

2.

 

Unplug the AC power cable from both the unit and the wall power socket. 

3.

 

Open the provided Transport Case by undoing all latches and removing the top cover. 

4.

 

Remove the instrument’s accessory foam insert from the Transport Case. 

5.

 

Carefully lift the instrument and place it into the Transport Case, within its dedicated custom 
foam cut out.  Note that there is only one proper orientation for the unit to be placed, otherwise 
the Transport Case will not close properly. 

6.

 

Replace the instrument’s accessory foam insert, and place all of the instrument’s accessories 
and power cable into their respective custom foam cut outs. 

7.

 

Replace the Transport Case cover and re-fasten all latches.   

8.

 

Lock or fasten each latch (where possible). 

9.

 

Where possible, two people shall move the unit (in its transport case) to its new location.  The 
Transport Case is also equipped with a retractable handle and wheels for single person 
transport. 

NOTE: 

Relocation within the same room is acceptable  to be completed without using the transport 
case; however, the transport case must be used otherwise.  Prior to shipment/relocation, ensure 
that the lid is fully and firmly attached to the  

main body of the case with all latches fastened 

or locked.  All parts, components and accessories required to operate the unit are must be 
included in the transport case.   

NOTE: 

The Sample Tray must be removed from the unit and placed in its proper location within the 
foam insert prior to transport. 

Summary of Contents for ChipCHECK CC120

Page 1: ...SIGNS DISCLOSED HEREIN ARE THE EXCLUSIVE PROPERTY OF GASTOPS LTD AND ARE NOT TO BE USED OR DISCLOSED TO ANY OTHERS WITHOUT THE WRITTEN CONSENT OF Gastops LTD ANY RECIPIENT BY ITS RETENTION AND USE AGR...

Page 2: ...unit requires service then it shall only be serviced or repaired by Gastops authorized technicians For the maintenance procedures discussed herein there is no need to remove any covers or modify the u...

Page 3: ...ng Once in Maintenance mode press the Conduct Maintenance button near the middle of the screen The Sample Receptacle will automatically displace further outside of the unit providing the required acce...

Page 4: ...n Figure 6 2 Maintenance Holder Receptacle Position Left Containment Window Assembly Right 6 1 2 Cleaning the Containment Window If the Containment Window is to be cleaned not replaced then the follow...

Page 5: ...wly drag the Isopropyl soaked cloth over the surface of the Containment Window pulling parallel to the exposed surface of the Containment Window ensuring no pressure is applied to the window itself Re...

Page 6: ...y Module Follow the set up instructions as described in Section 3 3 Figure 6 3 Power Entry Module and Main Power Switch 6 2 Unit Service No Allowable Service In the event of any equipment failure or c...

Page 7: ...designated customer support representative Refer to Section 8 0 Table 6 1 Basic Trouble Shooting Procedures Symptom Possible Remedy Software interface does not respond to user input Conduct a complet...

Page 8: ...on 5 4 The unit fails both Standardization and Calibration Power cycle the unit Perform the maintenance sequences defined in Section 6 0 Repeat Calibration and the complete Standardization If the unit...

Page 9: ...required Please call customer support Conduct a complete unit shutdown and power down Then power up and restart the unit as per Section 3 3 If this does not work then contact Gastops Customer Support...

Page 10: ...emovable foam insert protects the front face of the ChipCHECK unit When unboxing the unit this insert must be removed before the ChipCHECK can be removed from the transport case To remove the unit fro...

Page 11: ...d custom foam cut out Note that there is only one proper orientation for the unit to be placed otherwise the Transport Case will not close properly 6 Replace the instrument s accessory foam insert and...

Page 12: ...ions Foam insert removed Figure 7 4 Foam Insert Storage Locations Space for Operator s Manual located beneath ChipCHECK unit Storage slots beneath foam insert for Sample Cases Calibration Standardizat...

Page 13: ...200x Sample Patches 1x Standardization Standard 1x Calibration Standard 20X Sample Patch Cases 1x USB Ethernet Adapter Additional optional accessories may include the following External keyboard Exter...

Page 14: ...restrictions and conditions as set out in this publication 8 0 Customer Support Gastops technical support representatives for ChipCHECK can be reached at Phone 1 613 744 3530 Fax 1 613 744 8846 Toll...

Reviews: