background image

Gastops Ltd. 

 

 

C009423, Rev. 4 

 

page 45 

All use of this publication and ChipCHECK by user shall be subject to the restrictions and conditions as set out in this publication. 

 

Remove the Containment Window from its housing by carefully lifting the metallic cover 
located on top of the Sample Receptacle.  This cover is held in place by magnets.  As a result, 
a vertical lift shall remove the lid from the assembly.  Be sure to not apply much force to the 
Receptacle assembly. 

 

Place the Containment Window on a non-abrasive, lint free cloth/paper, with its bottom face 
pointing upwards (i.e.: the face of the Containment Window that is exposed to the 
contaminants placed in a manner that will allow for it to be cleaned). 

 

During this process ensure to not touch the surface of the Containment Window with bare 
fingers.  Gloves are required. 

 

Place one or two drops of Isopropyl Alcohol on the exposed face of the Containment 
Window. 

 

Gently place the non-abrasive, lint free cloth/paper onto the Isopropyl Alcohol and allow for 
a few seconds for the Isopropyl to propagate through the cloth. 

 

Carefully and slowly drag the Isopropyl soaked cloth over the surface of the Containment 
Window (pulling parallel to the exposed surface of the Containment Window ensuring no 
pressure is applied to the window itself). 

 

Repeat these steps as required. 

 

Once the Containment Window is clean, place it back onto the Sample Receptacle ensuring 
that the Containment Window is properly seated on the centering features of the 
Receptacle. 

 

During this process ensure to not touch the surface of the Containment Window with bare 
fingers.  Gloves are required. 

 

Carefully place the metallic cover back over the seated Containment Window – ensure that 
the cover is fully seated against the (note that the Containment Window metallic cover can 
only be assembled in one orientation). 

 

Lastly, return to the software interface and press the ‘Maintenance Complete’ button. 

 

The full Sample Receptacle will automatically re-insert itself back into the ejection position. 

 

Replace the Sample Tray, and return to the ‘Welcome’ screen for normal operation. 

6.1.3 Replacing the Fuse  

In the event that a fuse requires replacement, then the following procedure must be followed.   

WARNING:  

Disconnect supply before servicing (note that service/servicing can only be completed by 
Gastops authorized technicians).   

Disconnect supply before replacing fuse. 

Summary of Contents for ChipCHECK CC120

Page 1: ...SIGNS DISCLOSED HEREIN ARE THE EXCLUSIVE PROPERTY OF GASTOPS LTD AND ARE NOT TO BE USED OR DISCLOSED TO ANY OTHERS WITHOUT THE WRITTEN CONSENT OF Gastops LTD ANY RECIPIENT BY ITS RETENTION AND USE AGR...

Page 2: ...unit requires service then it shall only be serviced or repaired by Gastops authorized technicians For the maintenance procedures discussed herein there is no need to remove any covers or modify the u...

Page 3: ...ng Once in Maintenance mode press the Conduct Maintenance button near the middle of the screen The Sample Receptacle will automatically displace further outside of the unit providing the required acce...

Page 4: ...n Figure 6 2 Maintenance Holder Receptacle Position Left Containment Window Assembly Right 6 1 2 Cleaning the Containment Window If the Containment Window is to be cleaned not replaced then the follow...

Page 5: ...wly drag the Isopropyl soaked cloth over the surface of the Containment Window pulling parallel to the exposed surface of the Containment Window ensuring no pressure is applied to the window itself Re...

Page 6: ...y Module Follow the set up instructions as described in Section 3 3 Figure 6 3 Power Entry Module and Main Power Switch 6 2 Unit Service No Allowable Service In the event of any equipment failure or c...

Page 7: ...designated customer support representative Refer to Section 8 0 Table 6 1 Basic Trouble Shooting Procedures Symptom Possible Remedy Software interface does not respond to user input Conduct a complet...

Page 8: ...on 5 4 The unit fails both Standardization and Calibration Power cycle the unit Perform the maintenance sequences defined in Section 6 0 Repeat Calibration and the complete Standardization If the unit...

Page 9: ...required Please call customer support Conduct a complete unit shutdown and power down Then power up and restart the unit as per Section 3 3 If this does not work then contact Gastops Customer Support...

Page 10: ...emovable foam insert protects the front face of the ChipCHECK unit When unboxing the unit this insert must be removed before the ChipCHECK can be removed from the transport case To remove the unit fro...

Page 11: ...d custom foam cut out Note that there is only one proper orientation for the unit to be placed otherwise the Transport Case will not close properly 6 Replace the instrument s accessory foam insert and...

Page 12: ...ions Foam insert removed Figure 7 4 Foam Insert Storage Locations Space for Operator s Manual located beneath ChipCHECK unit Storage slots beneath foam insert for Sample Cases Calibration Standardizat...

Page 13: ...200x Sample Patches 1x Standardization Standard 1x Calibration Standard 20X Sample Patch Cases 1x USB Ethernet Adapter Additional optional accessories may include the following External keyboard Exter...

Page 14: ...restrictions and conditions as set out in this publication 8 0 Customer Support Gastops technical support representatives for ChipCHECK can be reached at Phone 1 613 744 3530 Fax 1 613 744 8846 Toll...

Reviews: