2.35  Pump Placement
 

Rubber feet are attached to the pump.  Rubber feet help to isolate noise and eliminate    

 

 

creeping.  All Aspiration/Filtration Vacuum Systems should be placed on a clea, horizontal  

 

 

surface.  Do not block the ventilation holes located on the case.

WARNING

 

 

 

Always vent pump exhaust to a hood or dilute exhaust gases 

 

 

 

appropriately when pumping potentially hazardous fumes.

2.40  Application Connection

2.41  Vacuum ports

 

Vacuum ports and other GEMINI

 pump components are identifi ed in Figure 1.  Be certain to  

 

 

connect vacuum application to the vacuum pump inlet.  See below identifi cation signs 

for inlet  

 

 

and exhaust ports.

 

 

 

INLET SIGN   

 

 

 

EXHAUST SIGN

CAUTION

 

 

 

Never plug or impede the vacuum pump exhaust port.  

 

 

 

Pressure build-up in the pump will cause pump to stall.

2.42  Tubing Connection
 

Use non-collapsible tubing for all vacuum connections.  Connect to pump inlet using 3/8”  

 

 

(10mm) ID tubing, keeping tubing length as short as possible.  Tubing should be chemically  

 

 

resistant to any vapors potentially evolved by the vacuum application.

 
2.43  Traps

 

GEMINI

 pumps are designed to work with aqueous vapors and select organic or acid/base   

 

fumes.  Always utilize a trap to minimize chemical fume contact with pump surfaces.  Empty  

 

 

the trap as needed to prevent fl uid ingestion into the vacuum pump.

CAUTION

 

 

 

Always check pump compatibility to potential application 

 

 

 

fumes prior to using your GEMINI

 pump with acid, base, 

 

 

 

or organic vapors / gases.  Serious damage to the pump 

 

 

 

can result from use with fumes that chemically compromise 

 

 

 

pump contact surfaces.

WARNING

 

 

 

Pumping fl ammable vapors or gases can result in fi re or 

 

 

 

explosion.

6

Summary of Contents for WELCH GEMINI 2050

Page 1: ...mited by compatibility with pump materials of construction Always verify vapor chemical compatibility prior to pump use and utilize proper cleaning methods to remove chemical residuals from the pump a...

Page 2: ...ction could result in injury or death to people CAUTION Cautions are found where failure to observe the instruction could result in damage to the equipment associated equipment and process These units...

Page 3: ...02 INSTALLATION 2 10 Introduction 2 20 Unpacking 2 30 Pump Location 2 40 Application Connection Section 03 OPERATION 3 10 Starting Procedures Applications Quali cation 3 20 Leak Detection 3 30 Pump Cl...

Page 4: ...part of the body 1 18 Use only in well ventilated areas 1 19 Never block pump exhaust Serious injury can result from pressure build up 1 20 To Reduce Risk Of Electrical Shock 1 21 Do not disassemble...

Page 5: ...urrounding temperature remains between 10 C and 40 C 50 F and 104 F Always check to insure the location chosen is protected from direct or indirect moisture contact 2 32 Electrical Source GEMINI pumps...

Page 6: ...port Pressure build up in the pump will cause pump to stall 2 42 Tubing Connection Use non collapsible tubing for all vacuum connections Connect to pump inlet using 3 8 10mm ID tubing keeping tubing...

Page 7: ...ons as well as certain alcohols ethanol isopropanol and carboxylic acids acetic acid It is important to check for pump material compatibility with potential fumes that can be drawn off your vacuum app...

Page 8: ...mBar GEMINI Start Up After verifying that your vacuum application is correctly con gured per Section 2 Installation turn on the pump On Off switch Welch recommends running the pump for a few minutes...

Page 9: ...r vapors through the pump for 5 minutes to clean pump contact surfaces GEMINI System Shutdown Procedures After use Welch recommends the pump be run for about 2 minutes disconnected from the vacuum pro...

Page 10: ...ower Failure If pump does not turn on the power adapter can be replaced Contact your Welch representative for information 10 SECTION 6 SPECIFICATION 6 10 SPECIFICATION TABLE 5 10 5 11 5 20 5 21 Parame...

Page 11: ...for any period however short will be repaired under this warranty rather than replaced The warranty is effective for one year from the date of original purchase when 1 The warranty card has been comp...

Page 12: ...ca Gardner Denver Thomas Inc Welch Vacuum Technology 5621 W Howard St Niles IL 60714 USA Phone 847 676 8800 Fax 847 677 8606 Email gdwelchvacuum gardnerdenver com Asia Rietschle Thomas Asia Paci c Ltd...

Reviews: